Authors, reviews, migrates, simulates, and troubleshoots official Opentrons Python Protocol API v2 protocols for Flex and OT-2 robots. Use for robot-specific liquid handling, deck and labware setup, pipettes, modules, runtime parameters, liquid classes, and Opentrons App analysis. Use pylabrobot instead when one workflow must support multiple robot vendors.

references/modules_and_deck.md

View on GitHub
# Modules and Deck Guide

Module compatibility is both an API question and a physical hardware question.
Confirm the robot model, module generation, caddy or adapter, firmware, deck
location, and labware before authoring commands.

## Robot Compatibility

| Hardware | Flex | OT-2 | Protocol API load name |
| --- | --- | --- | --- |
| Absorbance Plate Reader | Yes | No | `absorbanceReaderV1` |
| Flex Stacker | Yes | No | `flexStackerModuleV1` |
| Heater-Shaker GEN1 | Yes | Yes | `heaterShakerModuleV1` |
| Magnetic Block GEN1 | Yes | No | `magneticBlockV1` |
| Magnetic Module GEN1/GEN2 | No | Yes | `magnetic module` / `magnetic module gen2` |
| Temperature Module GEN2 | Yes | Yes | `temperature module gen2` |
| Thermocycler GEN2 | Yes | Yes | `thermocyclerModuleV2` |

Notes:

- Flex uses the passive Magnetic Block. The powered OT-2 Magnetic Module is not
  supported on Flex.
- Flex requires Thermocycler GEN2 for gripper-compatible operation.
- Older module generations may be supported only on OT-2. Match the load name
  to the exact hardware label.
- The HEPA/UV accessory is part of Flex operations but is not loaded and
  controlled as a Protocol API module.

See the current
[Flex module list](https://docs.opentrons.com/flex/modules/) and
[OT-2 module list](https://docs.opentrons.com/ot-2/modules/) before relying on
this snapshot.

## Deck Models

### Flex

- Working deck: A1–D3.
- Staging area: A4–D4; pipettes cannot reach it, but the Gripper can.
- Trash bins are loaded explicitly in supported column-1 or column-3 slots.
- The waste chute is loaded with `protocol.load_waste_chute()` and occupies D3.
- Powered module caddies and staging slots can conflict in the same row.
- Column-4 hardware can reserve or move through the corresponding column-3
  position.

```python
trash = protocol.load_trash_bin("A3")
chute = protocol.load_waste_chute()  # D3 only; do not load both into D3
```

Load only the fixtures physically installed for the run.

### OT-2

- User deck slots: 1–11.
- Fixed trash: slot 12.
- Do not call `load_trash_bin()` for the fixed trash.
- No Gripper or staging area.
- Labware moves are manual operator actions.

## Module Loading

```python
heater_shaker = protocol.load_module(
    module_name="heaterShakerModuleV1",
    location="D1",
)
```

Loading a powered module makes it a run requirement. The App will prevent the
run from starting unless the expected connected module is available.

Load labware through the module:

```python
hs_adapter = heater_shaker.load_adapter("opentrons_96_flat_bottom_adapter")
plate = hs_adapter.load_labware(
    "nest_96_wellplate_200ul_flat",
    label="Mixing Plate",
)
```

For a separate adapter:

```python
temperature_module = protocol.load_module(
    "temperature module gen2",
    "D3",
)
adapter = temperature_module.load_adapter(
    "opentrons_96_well_aluminum_block"
)
plate = adapter.load_labware(
    "opentrons_96_wellplate_200ul_pcr_full_skirt"
)
```

The API generally cannot prove that every unusual module, adapter, and labware
combination is physically valid. Check the official compatibility list.

## Temperature Module

```python
temperature_module.set_temperature(celsius=4)
# Pipetting or incubation commands...
temperature_module.deactivate()
```

`set_temperature()` blocks until the target is reached. In API 2.27+,
`start_set_temperature()` can run concurrently and returns a task. Wait for the
task before relying on the target temperature.

Operational checks:

- Use the correct aluminum block or adapter.
- Include condensation and cold-surface effects in the wet-lab method.
- Deactivate when the method no longer requires control.
- Do not assume the liquid itself instantly reaches the module temperature.

## Heater-Shaker

Typical sequence:

```python
heater_shaker.close_labware_latch()
heater_shaker.set_and_wait_for_temperature(37)
heater_shaker.set_and_wait_for_shake_speed(1000)
protocol.delay(minutes=5)
heater_shaker.deactivate_shaker()
heater_shaker.deactivate_heater()
```

Current context methods include `set_target_temperature()`,
`set_and_wait_for_temperature()`, `wait_for_temperature()`,
`set_and_wait_for_shake_speed()`, `deactivate_shaker()`, and
`deactivate_heater()`.

Always:

- Close the latch before shaking.
- Stop shaking before opening the latch or moving labware.
- Check maximum speed for the exact labware and fill volume.
- Account for clearance restrictions in neighboring slots.

API 2.27+ provides nonblocking shake and temperature operations for concurrent
work. Keep task handles and wait before a dependent step.

## Thermocycler

```python
thermocycler = protocol.load_module("thermocyclerModuleV2")
plate = thermocycler.load_labware(
    "opentrons_96_wellplate_200ul_pcr_full_skirt"
)

thermocycler.open_lid()
# Pipette into plate.
thermocycler.close_lid()
thermocycler.set_lid_temperature(temperature=105)
thermocycler.set_block_temperature(
    temperature=95,
    hold_time_seconds=180,
    block_max_volume=25,
)
thermocycler.execute_profile(
    steps=[
        {"temperature": 95, "hold_time_seconds": 15},
        {"temperature": 60, "hold_time_seconds": 30},
        {"temperature": 72, "hold_time_seconds": 30},
    ],
    repetitions=35,
    block_max_volume=25,
)
thermocycler.set_block_temperature(
    temperature=4,
    block_max_volume=25,
)
thermocycler.deactivate_lid()
```

API 2.28 adds an optional `ramp_rate` to block-temperature commands.

Checks:

- Correct plate and seal for the thermocycler.
- Before cycling, pause to confirm the compatible sealing setup. A reusable
  GEN2 automation seal attaches to the module lid, while a disposable Tough
  auto-sealing lid covers a compatible PCR plate. Never combine those two;
  follow the [lid-seal instructions](https://docs.opentrons.com/thermocycler/lid-seals/).
  Closing the heated lid does not itself establish correct seal installation.
- Lid closed before heating and cycling.
- `block_max_volume` matches the actual per-well reaction volume.
- Final hold behavior is intentional.
- Lid and block are deactivated or left active according to the method.
- No deck item conflicts with the thermocycler footprint.

## Magnetic Block, Flex

The Magnetic Block is passive. It has no engage or disengage methods.
Separation happens by moving compatible labware onto and off the block:

```python
magnetic_block = protocol.load_module("magneticBlockV1", "D1")

protocol.move_labware(
    labware=plate,
    new_location=magnetic_block,
    use_gripper=True,
)
protocol.delay(minutes=5)
protocol.move_labware(
    labware=plate,
    new_location="C1",
    use_gripper=True,
)
```

Confirm gripper compatibility and plate orientation. Avoid aspirating beads by
validating settle time, aspiration side, bottom clearance, and flow rate.

## Magnetic Module, OT-2

The powered Magnetic Module is OT-2-only:

```python
magnetic_module = protocol.load_module(
    "magnetic module gen2",
    "1",
)
plate = magnetic_module.load_labware(
    "nest_96_wellplate_2ml_deep"
)

magnetic_module.engage(height_from_base=6.5)
protocol.delay(minutes=5)
magnetic_module.disengage()
```

Engage height is labware- and assay-specific. Do not copy an arbitrary height
from another plate. The Magnetic Module is discontinued but remains supported
for existing OT-2 hardware.

## Absorbance Plate Reader, Flex API 2.21+

The reader is Flex-only and loads in A3–D3. Its caddy also uses the
corresponding column-4 location for lid travel.

Required workflow:

1. Load the module.
2. Close the lid with no plate inside.
3. Initialize the reader.
4. Open the lid.
5. Move a compatible plate onto the module with the Gripper.
6. Close the lid.
7. Read.

```python
reader = protocol.load_module(
    module_name="absorbanceReaderV1",
    location="D3",
)
plate = protocol.load_labware(
    "corning_96_wellplate_360ul_flat",
    "C2",
)

reader.close_lid()
reader.initialize(
    mode="multi",
    wavelengths=[450, 562, 600],
)
reader.open_lid()
protocol.move_labware(
    labware=plate,
    new_location=reader,
    use_gripper=True,
)
reader.close_lid()
data = reader.read(export_filename="plate_data")
```

Default hardware wavelengths are 450, 562, 600, and 650 nm.

`read()` returns:

```python
dict[int, dict[str, float]]
```

The first key is wavelength; the second is well name. In simulation every
measurement is zero, and no output file is written. Guard calculations that
would divide by a reading:

```python
if not protocol.is_simulating():
    normalized = data[450]["A1"] / data[450]["H12"]
```

Do not move the plate-reader lid manually.

## Flex Stacker, API 2.25+

Up to four Stackers attach to the right side of Flex. Each shuttle is addressed
as a column-4 location:

```python
stacker = protocol.load_module(
    module_name="flexStackerModuleV1",
    location="A4",
)
```

Configure one labware type per Stacker:

```python
stacker.set_stored_labware(
    load_name="opentrons_flex_96_tiprack_200ul",
    count=5,
    lid="opentrons_flex_tiprack_lid",
)
```

Retrieve and move:

```python
tip_rack = stacker.retrieve()
protocol.move_labware(
    labware=tip_rack,
    new_location="B2",
    use_gripper=True,
)
```

Return the same configured type to storage (with its required lid restored):

```python
protocol.move_labware(
    labware=tip_rack,
    new_location=stacker,
    use_gripper=True,
)
stacker.store()
```

`fill()` and `empty()` pause for manual loading or unloading.

Constraints:

- Configure stored labware before `retrieve()` or `store()`.
- Only one labware type per Stacker at a time.
- Flex tip racks need compatible lids to stack.
- Remove the tip-rack lid with a validated lid move before pipetting; restore
  it before storage. A retrieved lidded rack is not yet ready for tip pickup.
- The Stacker does not identify what an operator physically loaded.
- Reserve the corresponding row's shuttle path and check column-3 conflicts.
- Use capacity helper methods for the exact labware height.

## Moving Labware and Lids

Flex automated move:

```python
protocol.move_labware(
    labware=plate,
    new_location="C2",
    use_gripper=True,
)
```

Manual move:

```python
protocol.move_labware(
    labware=plate,
    new_location="C2",
    use_gripper=False,
)
```

A manual move pauses for the operator. Make sure the message and run setup make
the source and destination unambiguous.

API 2.23+ supports lid stacks and `move_lid()`. Confirm lid and labware
compatibility, orientation, stack quantity, and disposal location.

## Concurrent Module Actions, API 2.27+

Concurrent methods can overlap long module actions with independent pipetting:

- Temperature: `start_set_temperature()`.
- Heater-Shaker: `set_target_temperature()` and `set_shake_speed()`.
- Thermocycler: `start_set_block_temperature()`,
  `start_set_lid_temperature()`, and `start_execute_profile()`.

Pattern:

1. Start the operation and keep the returned task.
2. Perform only physically independent commands.
3. Wait for the task before any step that assumes completion.

```python
heat_task = heater_shaker.set_target_temperature(celsius=37)
# Independent work can run here.
protocol.wait_for_tasks([heat_task])
incubation = protocol.create_timer(seconds=60)
protocol.wait_for_tasks([incubation])
heater_shaker.deactivate_heater()
```

Use `set_target_temperature()` for the Heater-Shaker; it does not have the
Temperature Module's `start_set_temperature()` method. Wait through the protocol
context with `wait_for_tasks([task])`, not an invented `task.wait()` method.

Do not create concurrency merely to shorten runtime. Check deck access,
vibration, thermal dependencies, and collision risk.

## Module State Checklist

- Correct robot and module generation.
- Correct load name and API level.
- Valid deck slot and no footprint conflict.
- Correct caddy, adapter, and labware.
- Latch or lid state is safe before motion.
- Temperature, speed, and timing are validated for the method.
- Every started concurrent task is awaited.
- Gripper moves use compatible labware and clear paths.
- Modules are deactivated or intentionally held at protocol end.
- App analysis and a physical dry run cover the exact configuration.
planning-and-task-breakdownaddyosmani/agent-skillsBreaks work into ordered tasks. Use when you have a spec or clear requirements and need to break work into implementable tasks. Use when a task feels too large to start, when you need to estimate scope, or when parallel work is possible.Stars: 119kUpdated Claude CodeCodexVerifiedInstall
screenpipethedotmack/claude-memSearch screen and audio history recorded by Screenpipe and optionally save selected context into claude-mem. Use when the user asks to recall desktop activity, a meeting, or something seen in another app through Screenpipe.Stars: 118kUpdated Claude CodeCodexVerifiedInstall
contact-widgetnexu-io/open-designSelf-contained floating chat widget with welcome screen, social links, meeting button, and message input. Single HTML file, zero dependencies.Stars: 108kUpdated Claude CodeCodexVerifiedInstall
meeting-notesnexu-io/open-designMeeting notes page — title bar with attendees, agenda checklist, decisions block, action items table with owners + dates, and a "next meeting" footer. Use when the brief mentions "meeting notes", "minutes", "1:1 notes", "all-hands recap", or "会议纪要".Stars: 108kUpdated Claude CodeCodexVerifiedInstall